Opening of the detector
Loosen the 4mm hex screw (rep.1) locking the cover before removing the detector cover.
2.7.3
Cable preparation
The cable will be supplied from the controller at the measurement point. The passage, support,
and protection of the cable shall be done according to best practice.
2.7.4
Cable entry
It is essential to follow the instructions given by the manufacturer of the cable gland
and to connect the shielding properly. Cable-gland or adaptor shall be M25 x 1.5
and flameproof certified.
Detector
NPO60GB
Revision F.2
Figure 16: Locking screw of the cover
Detector
Shield
Washer
Seal
Shielding
Figure 17: non-armoured cable gland type
Seal
Armour grounding
Figure 18: armoured cable gland type
FIXED POINTGAS MONITOR
Non-armoured cable
Cable clamp
Seal
Armoured cable
OLCT 60
USER MANUAL