Download Print this page

Emerson Rosemount 2521 Quick Start Manual page 11

Solids level switch, vibrating fork
Hide thumbs Also See for Rosemount 2521:

Advertisement

October 2020
2.1.2
Solids in Water
The detection of solids in water is supported by the Rosemount 2521S only.
An installation example can be seen in
2.1.3
Mechanical load
The load at the mounting point must not exceed 300 Nm (Rosemount 2521
with an extended length fork).
Figure 2-1: Maximum Mechanical Load
A
A. Mounting point
B. Mechanical load
2.1.4
Vertical installations
Table 2-1
maximum deviations from a normal vertical installation.
Table 2-1: Maximum Vertical Deviation
Maximum deviation
45°
> 45°
Quick Start Guide
B
provides the maximum fork lengths and the corresponding
Figure
1-2.
Maximum fork length
157.5 in. (4000 mm)
47.24 in. (1200 mm)
23.62 in. (600 mm)
Quick Start Guide
11

Advertisement

loading