PlasmaQuant MS product family
9.4
Detector calibration
Set
ATTENUATION
Set the
DWELL TIME
Fig. 50
Configuring scan settings for attenuation calibration
3. Open the S
tab.
AMPLING
4. Set appropriate sample introduction parameters, such as sample uptake delay,
spray chamber temperature, and autosampler settings.
5. Save the worksheet by clicking the [S
6. Exit the E
M
DIT
ETHOD
1. Select the D
A
ETECTOR
2. Select the detector calibration worksheet by clicking [B
3. Set up the required number of calibrations. Five calibrations are recommended.
4. Click the [S
] button to begin a detector calibration.
TART
5. Present required solution A and/or solution B at the prompt.
to H
for all isotopes.
IGH
to at least 50,000 µs.
] button.
AVE
window.
tab from the I
TTENUATION
Detector calibration
window.
NSTRUMENT
].
ROWSE
91