Omicron Lab Bode 100 Detailed Functional & Calibration Check page 9

Vector network analyzer
Hide thumbs Also See for Bode 100:
Table of Contents

Advertisement

Figure 21: gain/phase accuracy - trace settings
Expected Measurement results:
Gain:
0 dB
Tolerance:
max. ± 0.2 dB from 1 Hz to 50 MHz
Phase:
Tolerance:
max ± 3° from 1 Hz to 50 MHz
If you measure a higher gain/phase deviation please send in the Bode 100 for calibration /
adjustment.
The following measurement results are an example and can be slightly different from the
measurement with other devices.
Figure 22: gain/ phase accuracy – magnitude (dB) measurement
Smart Measurement Solutions
Smart Measurement Solutions
Detailed Functional & Calibration Check of Bode 100
®