Omicron Lab Bode 100 Detailed Functional & Calibration Check page 5

Vector network analyzer
Hide thumbs Also See for Bode 100:
Table of Contents

Advertisement

Equipment Setup:
Please select the measurement type "Gain / Phase", select the source settings as follows and start a
continuous measurement:
Frequency:
10 MHz - Fixed
Output level: 0 dBm
Figure 7: output frequency accuracy - measurement type setting
Figure 8: output frequency accuracy - measurement setting
Figure 9: output frequency accuracy - start continuous measurement
Expected Measurement result:
Measured frequency: 10 MHz
Maximum tolerance for Bode 100 Revision 1:
± 15 ppm ≙ ± 150 Hz @ 10 MHz (< 1 year after adjustment)
± 25 ppm ≙ ± 250 Hz @ 10 MHz (< 3 years after adjustment)
Maximum tolerance for Bode 100 Revision 2:
± 2 ppm ≙ ± 20 Hz @ 10 MHz (< 1 year after adjustment)
± 4 ppm ≙ ± 40 Hz @ 10 MHz (< 3 years after adjustment)
If you measure a higher frequency deviation, please send the Bode 100 in for calibration / adjustment.