Bode 100 - Information
Detailed Functional & Calibration Check of Bode 100
Page 13 of 21
Figure 30: impedance accuracy - measurement & trace settings
Expected Measurement results:
Reflection:
max. - 35 dB for 1 Hz to 50 MHz
Figure 31: impedance accuracy – reflection magnitude (dB) measurement
If you measure a higher reflection than -35 dB please re-start the Internal Device Calibration and re-
perform this test. If this does not help, please send the Bode 100 in for calibration / adjustment.
®
Smart Measurement Solutions
Smart Measurement Solutions