Omicron Lab Bode 100 Detailed Functional & Calibration Check page 17

Vector network analyzer
Hide thumbs Also See for Bode 100:
Table of Contents

Advertisement

Then start a single measurement for each attenuator setting as seen below.
ATTENTION:
The Level needs to be -10 dBm to avoid an overload at the input of Bode 100.
Expected Measurement results:
Attenuator CH1
0 dB
10 dB
20 dB
30 dB
40 dB
0 dB
0 dB
0 dB
0 dB
10 dB
10 dB
10 dB
10 dB
20 dB
20 dB
20 dB
20 dB
30 dB
30 dB
30 dB
30 dB
40 dB
40 dB
40 dB
40 dB
If you measure a higher gain/phase deviation, please send the Bode 100 in for calibration.
Smart Measurement Solutions
Smart Measurement Solutions
Attenuator CH2
0 dB
10 dB
20 dB
30 dB
40 dB
10 dB
20 dB
30 dB
40 dB
0 dB
20 dB
30 dB
40 dB
0 dB
10 dB
30 dB
40 dB
0 dB
10 dB
20 dB
40 dB
0 dB
10 dB
20 dB
30 dB
®
Detailed Functional & Calibration Check of Bode 100
Expected Result
Gain: 0dB
Tolerance: max. ± 0.2 dB for 1 Hz to 50 MHz
Phase: 0°
Tolerance: max ± 3° for 1 Hz to 50 MHz
Gain: 0dB
Tolerance: max. ± 0.2 dB for 1 Hz to 50 MHz
Phase: 0°
Tolerance: max ± 4° for 1 Hz to 50 MHz
Gain: 0dB
Tolerance: max. ± 0.3 dB for 1 Hz to 50 MHz
Phase: 0°
Tolerance: max ± 6° for 1 Hz to 50 MHz
Bode 100 - Information
Page 17 of 21