After that, please start a continuous sweep.
Expected measurement results:
Receiver levels should look similar to the following when source frequency is set to 1 kHz:
OPEN
SHORT
LOAD (50 Ω)
In the following, the source frequency must be changed between 1 kHz and 100 kHz.
1 kHz
OPEN
>85 kΩ
<20 mΩ
SHORT
LOAD (50 Ω)
<51 Ω & >49 Ω (typ. 50 Ω)

2 Calibration service

ISO9001 Factory Calibration
Calibration and adjustment of the Bode 100 is performed in our ISO 9001-2008 certified development
center. If a unit is sent in for calibration, we perform a complete factory calibration including
adjustment like during manufacturing. This ensures that the Bode 100 has the same accuracy as
when it was manufactured.
IEC17025 Calibration
Together with an external calibration lab we offer an ISO/IEC17025 calibration as an add-on to our
factory calibration.
For further information please contact our support team:
Smart Measurement Solutions
Smart Measurement Solutions
Detailed Functional & Calibration Check of Bode 100
(typ. 100 kΩ)
(typ. 10 mΩ)
®
100 kHz
(typ. 90 kΩ)
>75 kΩ
<50 mΩ
(typ. 10 mΩ)
<51 Ω & >49 Ω (typ. 50 Ω)