Check Of Internal Connections & Output Impedance - Omicron Lab Bode 100 Detailed Functional & Calibration Check

Vector network analyzer
Hide thumbs Also See for Bode 100:
Table of Contents

Advertisement

1.10 Check of internal connections & output impedance
In the following, the internal connections from the receivers to the 50 Ω output impedance are tested.
Additionally, the output of the Bode 100 will be checked.
Measurement Setup:
OPEN:
SHORT:
LOAD
(50 Ω):
Equipment Setup:
Figure 41: Internal connection check - measurement mode
Use the "One-Port" impedance measurement method and set the frequency to "Fixed" as well as
follows:
Figure 42: Internal connection check - measurement settings