Tenma 72-17175 Manual page 3

Voltage/current calibrator

Advertisement

ELECTRICAL SPECIFICATIONS
Current Source
Range
0-19.99mA
0-24mA
* Output 0 to 24mA current for loads up to 500ohms
* Output >20mA current for loads up to 400ohms
* FS: full scale
Current Measurement
Range
0-19.99mA
0-24mA
* FS: full scale
Power and Current Measurement of Two Wire Loop
Range
0-19.99mA
0-24mA
* Provides power DC 12V ±2V to the loop and measures current.
* FS: full scale
DC mV source
Range
-199.9mV to
+199.9mV
* Output measured load impedance should >1K ohm.
* FS: full scale.
Note: the above specifications are tested under the environment RF field strength less
than 3V/M and frequency less than 30MHz.
MEASURING PROCEDURE
Current Source
Insert the black test lead into the negative (COM) jack and the red test lead into the
positive (mV-mA) jack.
Set the function switch to current (mA source) position.
Press the 0-19.9mA button to set the 0.01mA display resolution (max display is
19.99mA) or press the 0-24mA button to set the 0.1mA display resolution.
Adjust the calibration rotary control (CAL ADJUST) to generate the current output.
Current Measurement
Insert the black test lead into the negative (COM) jack and the red test lead into the
positive (mV-mA) jack.
Set the function switch to current (mA measure) position.
Press the 0-19.9mA button to set the 0.01mA display resolution (max display is
19.99mA) or press the 0-24mA button to set the 0.1mA display resolution.
Display resolution
0.01mA
0.1mA
Display resolution
0.01mA
0.1mA
Display resolution
0.01mA
0.1mA
Display resolution
0.1mV
Accuracy
±(0.25%FS+1d)
±(0.5%FS+1d)
Accuracy
±(0.25%FS+1d)
±(0.5%FS+1d)
Accuracy
±(0.25%FS+1d)
±(0.5%FS+1d)
Accuracy
±(0.25%FS+1d)