Miniox ® Responder Oxygen Detector; (P/Ns 710965 And 10008676); Instructions For Use And Maintenance; Warnings - MSA MiniOX 710965 Instruction Manual

Table of Contents

Advertisement

Section 3
®
MiniOX
Responder Oxygen
Detector

(P/Ns 710965 and 10008676)

Instructions for Use and Maintenance

" WARNINGS

FAILURE TO FOLLOW CAN RESULT IN
SERIOUS PERSONAL INJURY OR DEATH.
1. Alarm functions must be checked and a
response check must be performed daily.
If the instrument fails the response check
and cannot be calibrated, either the
sensor or battery must be replaced, or the
instrument must be serviced.
2. Calibration must be checked if the sensor or
battery is replaced, or if the instrument is
dropped or subjected to severe physical
shock. Make sure the instrument is in fresh
air when performing an FAS or calibration.
3. The oxygen sensor is a sealed unit
containing sulfuric acid electrolyte. If a
sensor develops a leak, dispose of it
properly. Should contact occur with skin
or clothing, rinse area with large
quantities of water. In case of eye contact,
immediately flush eyes for at least 15
minutes, holding eyes open; call
a physician.
4. Do not change battery in hazardous
locations.
5. Substitution of components may impair
intrinsic safety.
6. High or low pressure samples will give
erroneous readings; calibrate at altitude
of use.
7. Once a high alarm condition is detected
and the alarms latch, the displayed
reading may not fall within the specified
accuracy range for the instrument.