Installing/Replacing the Battery

Remove belt clip by pressing the two tabs near
the top and releasing from the back. Remove
the two slotted head screws on the back of the
case, located under the clip. Remove the 9-V
battery from the snap connector and replace
with a fresh 9-V battery or install a fresh 9-V
battery for first-time use. Replace the battery
cover and the two slotted screws; snap the belt
clip back. Wait one minute before turning the
instrument ON and zeroing it. If the instrument
has been out of service or the battery has been
out of the instrument for more than one day, it
is recommended that the instrument not be used
or zeroed for a period of 12 hours. This allows
the sensor time to equilibrate. The instrument
will not turn ON for 45 seconds. The backlight is
ON and --- is displayed during this period.

Replacing Sensor Assembly

Remove phillips screw from sensor assembly.
o
Rotate approximately 15
clockwise. Pull assembly
out. Remove the shorting bar from the new
assembly. Place the assembly in the instrument
o
and rotate approximately 15
counterclockwise;
insert the phillips screw and tighten.
It is recommended that the instrument not be
used or calibrated for a period of 12 hours once
the sensor has been installed. This allows the
sensor time to equilibrate.
6-5