has been out of service or the battery has been
out of the instrument for more than one day, it
is recommended that the instrument not be used
or zeroed for a period of 12 hours. This allows
the sensor time to equilibrate. The instrument
will not turn ON for 45 seconds. The backlight is
ON and --- is displayed during this period.

Replacing Sensor Assembly

Remove phillips screw from sensor assembly.
Rotate approximately 15
out. Remove the shorting bar from the new
assembly. Place the assembly in the instrument
and rotate approximately 15
insert the phillips screw and tighten.
It is recommended that the instrument not be
used or calibrated for a period of 12 hours once
the sensor has been installed. This allows the
sensor time to equilibrate.

Specifications

RANGE
RESPONSE TIME
TEMPERATURE
RANGE
(COMPENSATED)
HUMIDITY LIMITS
BATTERY LIFE
WARM-UP TIME
ALARM RANGE
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Section 5, MiniCl
Responder
2
o
clockwise. Pull assembly
o
counterclockwise;
0-15 ppm,
over range indicated by - - -
90 seconds or less
to 90% of final reading
o
o
o
o
-10
C to 40
C (14
F to 104
F)
o
o
40
C to 50
C if calibrated at
temperature of use and
recalibrated again if temperature
drops within the compensated
range
15 - 90% (non-condensing)
60 days with five minutes of
alarm per day
90 days with no alarm
As much as 12 hours when a
new sensor or battery is installed
0.2 to 4.9 ppm
5-5