To save batteries, you may enable the <Autoff> function in the menu (see
11.2.1).
chapter
• If the scales is not used for a prolonged period of time, remove the batter-
ies and store them separately. Any leaking electrolyte could damage the
scales.

7.6 Rechargeable battery operation (option)

The rechargeable battery is charged using the supplied power cord.
Before first use, charge the battery for at least 15 hours using the power cord.
To save the rechargeable battery, you may enable the <Autoff> function in the menu
11.2.1).
(see chapter
Once the rechargeable battery is discharged, the <instab> symbol is displayed.
To charge the battery, connect the power cord as soon as possible. The charging
time until the full charging status is reached is ca. 10 hours.

7.7 Connecting peripherals

Before you connect or disconnect any extra devices (printer, computer) to/from the
data interface, the scales should always be disconnected from the mains.
Use solely accessories and peripherals supplied by KERN with the scales, being per-
fectly compatible with it.

7.8 First start

To get accurate weighing results using electronic scales, ensure the scales achieves
the appropriate operating temperature (see "Heating time", chapter 1). During the
heating time, the scales must be connected to the power source (the socket, re-
chargeable battery or batteries).
The scales accuracy depends on the local standard gravity.
Always follow the guidelines in the "Adjustment" chapter.

7.9 Adjustment

As the standard gravity value is not the same in every spot on Earth, every display
with the scales plate connected should be adjusted, in line with the weighing rules
resulting from the laws of physics, to the standard gravity in the scales location (pro-
vided the scales system has not already been subject to factory adjustment in its lo-
cation). Such an adjustment process should be carried out during the first start, fol-
lowing every location change and also in the case of any ambient temperature fluctu-
ations. To ensure achieving accurate measurement date, it is also recommended to
carry out regular display adjustment also in the weighing mode.
 How to do it, see chapter 11.2.2
14
CKE-BA-e-2031