Honeywell SmartLine SLN 700 Series User Manual page 33

Level transmitter, non-contact radar
Hide thumbs Also See for SmartLine SLN 700 Series:
Table of Contents

Advertisement

3.3 Sensor
Status
3.4 Measure status
Press "Enter" to select whether to reset the above values.
3.5 Peak Values
Press "Enter" to select whether to reset the above values.
3.6 Calibration Date
new date after they performed a calibration.
Revision 1
Press
key to enter the next diagnostics measurement status when
the LCD displays menu number 3.2.
Operation status of the sensor is as follows.
Sensor Status
T
DB
Volt
Service Time
T is current sensor temperature in 'C;
DB is current signal strength in Db;
Volt is current power input voltage;
Service Time is device accumulated operation time.
Max Volt/Min Volt/ Min Volt Time
Max Volt is maximum power input voltage since recent power
on;
Min Volt is minimum power input voltage since recent power
on;
Min Volt Time is the time when device received min voltage
Max Distance/Min Distance
Max Distance is the maximum distance during the
measurement;
Min Distance is the minimum distance during the measurement;
The default value is calibration date from factory. User can enter the
SLN 700 SmartLine NCR Level Transmitter User's Manual
Page 25