Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Locks the keypad.
Locks or unlocks the front panel depending on the value
assigned to x:
0 - unlock front panel
1 - lock front panel
Reads the counter that tracks valve turns.
Sets the counter that tracks valve turns. Replace x with a
number from 0 to 99999.
For example, the command 'M0' sets the number to 0; you
use this command after installing a new valve.
Changes the valve position display to numbers and moves
the valve to the position n, where n is a number between 0
and 8.
Reads the counter that tracks the number of hours that
either the VALVEMATE or a valve has been in use.
Sets the counter that tracks the number of hours that either
the VALVEMATE or a valve has been in use. Replace x
with a number from 0 to 99999.
For example, the command 'N0' sets the number to 0; you
can use this command after installing a new valve.
Identifies the current valve position. Returns 'sp', where s is
status and p is position.
s - 'P' for Parked, 'M' for Moving, or 'E' for Error
p - 'A' to 'H'.
For example, if the valve was in position A, the
VALVEMATE returns 'PA'.
Changes the valve position display to letters and moves the
valve to the position set by x, where x is a letter between A
and H.
For example, if you want to go to position A, send the
command PA.