Installation
Rheodyne 5011, 5012, 7060, or 9060 valve. To
verify that the valve is properly installed:
1
If the valve has a loop, remove the loop.
2
Fill a syringe.
3
Connect the syringe to the center port and
dispense liquid into the port.
4
Verify that the liquid exits port 1. If it does not,
remove the valve, orient the valve in its home
position, and install the valve again. (Refer to
pages
2-13
and 2-14.) Then repeat steps 1–4.
5
Press the up arrow key.
6
Dispense additional liquid into the center port.
7
Verify that the liquid exits port 2.
Rheodyne 5031, 5032, 7030, or 9030 valve. To
verify that the valve is properly installed:
1
If the valve has a loop, remove the loop.
2
Fill a syringe.
3
Connect the syringe to port 2 and dispense
liquid into the port.
4
Verify that the liquid exits port 1. If it does not,
remove the valve, orient the valve in its home
position, and install the valve again. (Refer to
pages
2-13
and 2-14.) Then repeat steps 1–4.
5
Press the up arrow key.
6
Dispense additional liquid into port 2.
7
Verify that the liquid exits port 3.