GSIOC Commands
There are two kinds of commands that you can
issue from the controller to the VALVEMATE.
• Buffered commands (B) send instructions to the
VALVEMATE. These commands are executed
one at a time. The VALVEMATE returns a
period (.) to indicate that it accepted the buffered
command.
• Immediate commands (I) request status
information from the VALVEMATE. These
commands are executed immediately, temporarily
interrupting other GSIOC commands in progress,
and return the requested information.
When issuing GSIOC commands to the
VALVEMATE, separate the time points by at least
0.03 minute (or 1.8 seconds). For example:
Time
Unit
Command
(min) ID
Type
0.10
35
Buffered
0.13
35
Buffered
C
Command
Description
2
Moves
valve to
position 2.
3
Moves
valve to
position 3.