SYSTEM 3000 OPERATIONS AND MAINTENANCE MANUAL
4.12
FUNCTION GENERATOR AND A SCOPE.
Function
Generator
With the function generator and the scope connected to the cable, monitor input on
the scope and input a 1 volt peak (2 volt peak to peak) sine wave of the specific
frequency of interest. Measure the peak output at the other end of the cable on the
scope through the 50 ohm termination.
Use the formula :
Example:
4-36
Cable
Assembly
dB = 20 log (V out / V in)
Info: If you use 1 volt peak for V in, V in = 1
At 750 kHz you get a reading of 160 mV
.16 v
log .16
x 20
-15.9 dB loss
TROUBLESHOOTING GUIDE
50 ohm terminator
L-3 Klein Associates, Inc.
Scope