Electrical Alignment; Preparations; Equipment Required; Precautions On Adjustments - Sony HXCU-FB80 Service Manual

4k/hd camera control unit
Hide thumbs Also See for HXCU-FB80:
Table of Contents

Advertisement

4-1.

Preparations

4-1-1.

Equipment Required

• SD waveform/vector monitor
• Analog HD waveform monitor
• 3G/12G-SDI compatible waveform monitor
• Oscilloscope: Tektronix TDS460A or equivalent
• Frequency counter: Advantest TR5821AK or equivalent
• Audio oscillator: Tektronix SG-5010 or equivalent
• Audio analyzer
Related equipment
• HD color camera HXC-FB80
• Remote control panel RCP-1500 series
4-1-2.

Precautions on Adjustments

• Warm up the unit for about 10 minutes before starting adjustments.
• All measuring equipment must have been calibrated.
"4-1-4. Initial Settings"
• Adjustments for HXC-FB80 must have been completed.
4-1-3.

Connection of Equipment

Tip
Transmittable Optoelectrical composite cable: CCFN-25/50/100
HD color camera
HXC-FB80
HXCU-FB80

Electrical Alignment

must have been completed.
Optoelectrical
composite cable
CCA-5 cable
(max. 50 m)
Remote control panel
RCP-1500 series
Section 4
4K/HD camera control unit
HXCU-FB80
Audio
oscillator
Signal
generator
REFERENCE
/DPR-386
Waveform
monitor
PIX/CN-3598
75 Ω
75 Ω
4-1