Technical Data - Meiko TopLine Instructions For Use Manual

Washer-disinfector
Table of Contents

Advertisement

5

Technical data

Protection de-
gree
Protection
class
Installation
conditions
Fresh water
supply line
On-site sup-
ply line
Water flow
pressure
Water flow
Boiler content
DT-NDV
Steam pres-
sure
DT-NDD
Steam flow
DT-NDD
Disinfection
performance
Heat emission For 10 programme cycles/h:
Noise emis-
sion
9617722
MEIKO TopLine 10/20/40: IP X4; TopLine 30: IP X5
Class I in accordance with VDE 0750 / IEC 601
Max. height 2000 m above NHN, 5–40°C with max. 80% rel-
ative humidity, exclusively for indoor use, degree of soiling 2,
overvoltage category 2, mains voltage tolerance ± 10%, min.
illuminance ambient light: 200 lx
Air gap group A, type A according to DIN 1988 /
DIN EN 1717
DN 15 (½ inch), cold water 5-25°C, hot water 45-60°C. From
a biological perspective, the fresh water must be of drinking
water quality. This also applies to processed water.
100–500 kPa, before the solenoid valve
18 l/min
2.1 l
10–50 kPa
Approx. 5 kg/h
A
-value can be adjusted from 60–3,000
0
Total 0.7 kW / sensitive 0.45 kW / latent 0.25 kW
Workplace noise emissions: LpA = 48.2 dB(A)
The emissions noise level at the workplace for the docu-
mented load has been determined on the basis of: DIN EN
ISO 11204; precision class 3, measurement reliability ± 5 dB.
Level at user location: 61.2 dB(A)
Level at 1 m distance from the appliance: 59.5 dB(A)
20 / 60

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents