10.4 Daily Cleaning; 10.5 De-Scaling - Meiko TopLine Instructions For Use Manual

Washer-disinfector
Table of Contents

Advertisement

MEIKO T opLi ne

10.4 Daily cleaning

CAUTION – Material damage due to incorrect cleaning
Cleaning of parts made of stainless steel with unsuitable cleaning agents,
care products and cleaning utensils leads to damage, deposits or discoloura-
tions on the appliance.
Never use aggressive cleaning or scouring agents.
Never use cleaning agents that contain hydrochloride acid or bleaches
based on chlorine.
Do not use cleaning utensils previously used to clean non-stainless steel.
If the washer-disinfector is not used on a daily basis, start the programme once
a day or let it run automatically with the time switch (see service instructions).
Clean the outside of the washer-disinfector, especially the door handle and
membrane key pad, once a day. Only clean stainless steel with detergents
suitable for stainless steel.
Make sure that detergents do not attack the material, do not form coatings
and do not discolour the material.
Never clean the washer-disinfector with water hose, high pressure cleaner or
steam cleaner.
Only clean wash chamber seal with water or soap.

10.5 De-scaling

CAUTION – Destruction of plastic parts and seals from residues of de-
scaling agent
Completely flush the descaling agent out of the wash chamber.
Lime deposits (discolouration) in the wash chamber should be sprayed with
a commercially available liquid descaler (e.g. based on 10 - 20% citric acid
or phosphoric acid).
Mechanical treatment with a brush or cloth is necessary. Do not use steel
brushes. Observe the manufacturer's safety instructions and take suitable
protective measures (protective gloves and safety eyeware, disposable
apron if necessary).
After a short application time (5 minutes), close the wash chamber door and
immediately start a cleaning programme. This rinses the wash chamber.
53 / 60
9617722