Agilent Technologies E1476A User Manual page 117

64-channel, 3-wire multiplexer module
Table of Contents

Advertisement

Code
Error Message
-211
Trigger ignored
-213
Init Ignored
-222
Data out of range
-224
Illegal parameter value
-240
Hardware error
-310
System error
1500
External trigger source
already allocated
1510
Trigger source
non-existent
2000
Invalid card number
2001
Invalid channel
number
2006
Command not
supported on this card
2008
Scan list not initialized
2009
Too many channels in
channel list
2010
Scan mode not
allowed on this card
2011
Empty channel list
2012
Invalid Channel Range
2017
Config error 17, Slot 0
functions disabled
2600
Function not
supported on this card
2601
Channel list required
116 E1476A Error Messages
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Table C-2. Multiplexer Error Messages
Trigger received when scan not enabled. Trigger received after scan
complete. Trigger too fast.
Attempting to execute an INIT command when a scan is already in
progress.
Parameter value is outside valid range.
Attempting to execute a command with a parameter not applicable to the
command.
Command failed due to a hardware problem.
Internal driver error. This error can result if an excessively long parameter
list is entered.
Assigning an external trigger source to a switchbox when the trigger
source has already been assigned to another switchbox.
Selected trigger source is not available on this platform (e.g., some
triggers are not available on E1300/E1301 VXI B-Size mainframes).
Addressing a module (card) in a switchbox that is not part of the
switchbox.
Attempting to address a channel of a module in a switchbox that is not
supported by the module (e.g., channel 99 of a multiplexer module).
Sending a command to a module (card) in a switchbox that is
unsupported by the module.
Executing a scan without the INIT command.
Attempting to address more channels than available in the switchbox.
The selected scanning mode is not allowed with this module or you have
misspelled the mode parameter (see SCAN:MODE command).
No valid channels are specified in the channel_list.
Invalid channel(s) specified in SCAN <channel_list> command.
Attempting to begin scanning when no valid channel list is defined.
Attempt to run a downloaded scan list with ARM:COUNt set to a value
other than 1. Applies to FET switches only.
Sending a command to a module (card) in a switchbox that is not
supported by the module or switchbox.
Sending a command requiring a channel_list without the channel_list.
Potential Cause(s)
Appendix C