STATus:OPERation:CONDition?

STATus:OPERation:ENABle

Parameters
Comments
Example
STATus:OPERation:ENABle?
Comments
Example
Chapter 4
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
STATus:OPERation:CONDition?
Operation Status Group. The state represents conditions which are part of the
instrument's operation. The "SWITCH" driver does not set bit 8 in this register (see
STATus:OPERation[:EVENt]?).
STATus:OPERation:ENABle <unmask>
recorded in the Event Register to send a summary bit to the Status Byte Register (bit
7). For multiplexer modules, when bit 8 in the Operation Status Register is set to 1
and that bit is enabled by STATus:OPERation:ENABle, bit 7 in the Status Register
is set to 1.
Name
<unmask>
Setting Bit 7 of the Status Byte Register:
of the Status Byte Register to 1 after bit 8 of the Operation Status Register is set to 1.
[ROUTe:]SCAN
Related Commands:
Enabling Operation Status Register Bit 8
67$723(5(1$% 
STATus:OPERation:ENABle?
Status Group) are unmasked.
Returns a decimal weighted value from 0 to 65,535 indicating which
Output Format:
bits are set to true.
Maximum Value Returned:
STAT:OPER:ENAB <unmask> command. However, the maximum decimal
weighted value used in this module is 256 (bit 8 set to true).
Query the Operation Status Enable Register
67$723(5(1$%"
returns the state of the Condition Register in the
sets an enable mask to allow events
Type
Range of Values
numeric
0 through 65,535
STATus:OPERation:ENABle 256 sets bit 7
Enable bit 8 of the Operation Status
Register to be reported to bit 7 (OPR) in the
Status Byte Register.
returns which bits in the Event Register (Operation
The value returned is the value set by the
Query the Operation Status Enable