Solvent Refill - JDS Uniphase CleanBlast FCL-P1005 User Manual

Advanced fiber end face cleaning system with fbp probe microscope
Table of Contents

Advertisement

solvent refIll

Note: Do not refill the solvent reservoir if the LOW SOLVENT light is not illuminated. Overfilling the tank can
prevent the system from pressurizing properly.
1.
Turn OFF the system and disconnect the power cord from the AC power source.
2.
Press the cap lock down and rotate (counter-
clockwise) the refill cap to expose the refill
ports.
Note: You may hear a 'hiss' of air as the tank
depressurizes.
3.
Attach the solvent refill cap assembly to the
solvent bottle.
You must select one solvent type and use the
same solvent for the lifetime of the system.
4.
Push-in to attach the 2 quick connect valves
to the two refill ports, and invert - hold the
bottle until the system's cleaning solvent
reservoir is full.
5.
Press the valve release down to disconnect
the bottle and rotate the refill cap to original
position and turn the system ON.
6.
Prime the system (see page 16).
mAINTENANCE SChEDuLE
Depending on the type of cleaning tip used, you should expect
the following number of cleaning cycles per fUll solvent
reservoir.
2.5 mm tip
9,500 cycles
1.25 mm tip
12,500 cycles
ribbon/MtP tip
9,500 cycles
20
4
User ManUal
2
Cap Lock
solvent refill Cap
Assembly
3
Cleaning solvent Bottle
(see page 25 for solvent types)
5
Valve
release