sYsteM statUs InDICators
SYSTEm STATuS INDICATORS
POWER / READY
This dual indicator will alert the user when there is AC
power connected to the system (red), and when the
nozzle safety mechanism is disengaged, allowing the
cleaning cycle to activate (green).
Note: The run button on the handset will not
operate when this LED is red. To disengage the safety
mechanism, apply light pressure against the bulkhead
or patch cord and gently slide the handset forward.
When the POWER / READY LED turns GREEN, press the
RUN BUTTON on the handset to initiate cleaning.
LOW SOLvENT
This indicator will alert the user that the level of
cleaning solvent is too low for the system to operate
and will put the system into standby mode. The
solvent reservoir must be refilled before operation
can resume (see page 20 for solvent refill procedures).
The reservoir will hold 225 ml. (8 fl. oz.) of solvent,
enough for at least 8,000 cleaning cycles.
Note: Do not refill the solvent reservoir if the LOW
SOLVENT light is not illuminated. Overfilling the tank
can prevent the system from pressurizing properly.
SERvICE
This indicator will alert the user that the system
air filters will need to be replaced. When the reset
button is pressed, the system cycle counter will turn
to zero, and the system will count to 100,000 cycles
before alerting the next filter change (see page 21 for
air filter change procedures).
CleanBlast – PortaBle
Quick Connect
Valves
Air Filter
Ordering Part Number
• FCLP-FA-F1 (2 required per system)
solvent refill Cap
Assembly
Ordering Part Number
• FCLP-RCA-1
Cleaning solvent Bottle
Ordering Part Numbers
• FCLP-SOL1 (1 bottle)
• FCLP-SOL1-6 (6 bottles)