Clean - JDS Uniphase CleanBlast FCL-P1005 User Manual

Advanced fiber end face cleaning system with fbp probe microscope
Table of Contents

Advertisement

Clean: BULKHEAD
CleanBlast
Advanced Fiber End Face Cleaning System
1.
Install the appropriate cleaning tip
(fCLT-U25 shown) to the CleanBlast handset.
Note: See Appendix B: Cleaning Tips &
Adapters Guide on page 27.
PRImE
PRImE / PuRgE the nozzle to clear the solvent line if
you are using the system for the first time, after refill,
or if the system has been powered oFF for more than
4 hours.
A.
Push the prime button quickly 2 times.
B.
Aim the tip away from any surfaces or objects.
C.
Press the run button on the handset.
D.
Repeat steps A through C at least 2 more times.
2.
Insert the handset into the bulkhead and
push to disengage safety mechanism (the
POWER/READY light will turn green), and press
the run button to initiate cleaning.
3.
INSPECT
the bulkhead.
4.
Determine whether clean or dirty.
 If clean, do not touch it and CONNECT.
 If dirty, repeat
Note: Embedded/mated debris cannot be
cleaned, and bonded/burned debris must be
physically cleaned before using CleanBlast.
16

CLEAN.

1
Cleaning Tip
(FCLT-U25)
2
4
User ManUal
CLEAN
Prime Button
(press 2 times)
run Button
run Button
3

Hide quick links:

Advertisement

Table of Contents
loading