Ametek ORTEC 4002P Operating And Service Manual
Ametek ORTEC 4002P Operating And Service Manual

Ametek ORTEC 4002P Operating And Service Manual

Portable power supply

Advertisement

Quick Links

Model 4002P
Portable Power Supply
Operating and Service Manual
®
Printed in U.S.A.
ORTEC
Part No. 740310
0815
801 South Illinois Avenue
Manual Revision E
Oak Ridge, Tennessee 37830
United States of America

Advertisement

Table of Contents
loading

Summary of Contents for Ametek ORTEC 4002P

  • Page 1 Model 4002P Portable Power Supply Operating and Service Manual ® Printed in U.S.A. ORTEC Part No. 740310 0815 801 South Illinois Avenue Manual Revision E Oak Ridge, Tennessee 37830 United States of America...
  • Page 2 Advanced Measurement Technology, Inc. ® ® a/k/a/ ORTEC , a subsidiary of AMETEK , Inc. WARRANTY ORTEC* warrants that the items will be delivered free from defects in material or workmanship. ORTEC makes no other warranties, express or implied, and specifically NO WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
  • Page 3: Table Of Contents

    CONTENTS WARRANTY..............SAFETY INSTRUCTIONS AND SYMBOLS.
  • Page 4: Safety Instructions And Symbols

    SAFETY INSTRUCTIONS AND SYMBOLS This manual contains up to three levels of safety instructions that must be observed in order to avoid personal injury and/or damage to equipment or other property. These are: DANGER Indicates a hazard that could result in death or serious bodily harm if the safety instruction is not observed.
  • Page 5: Safety Warnings And Cleaning Instructions

    SAFETY WARNINGS AND CLEANING INSTRUCTIONS DANGER Opening the cover of this instrument is likely to expose dangerous voltages. Disconnect the instrument from all voltage sources while it is being opened. WARNING Using this instrument in a manner not specified by the manufacturer may impair the protection provided by the instrument.
  • Page 7: Description

    ORTEC MODEL 4002P PORTABLE POWER SUPPLY 1. DESCRIPTION connector block on the rear panel is used to change 1.1. GENERAL and display the selected input voltage. Also The ORTEC Model 4002P Portable Power Supply incorporated in this connector block are a fuse is ideally suited for providing power to preamplifiers, holder and the input power cord connector.
  • Page 8: Controls And Indicators

    thermal switch automatically turns off the power INPUT CURRENT Typically, 0.8 A rms with a supply. Under this thermal shut-down condition, both 48-W dc load and a 120-V ac input. Protected with a 2-A fuse on the 100- and 120-V ac settings, and the TEMP and the POWER lights are turned off.
  • Page 9: Electrical And Mechanical

    preamplifier power cables of most other NIM delivered to the rear-panel connectors.. manufacturers. Pin assignments are listed in Table 1. Each preamp power output connector can 2.5. ELECTRICAL AND MECHANICAL be expanded to three output connectors by using WEIGHT the optional Model 4002P-C1 Preamp Power Fan- Net 10.9 kg (24 lb) Out Cable.
  • Page 10: Selecting The Mains Voltage

    Disconnect the instrument from the ac power 3.4. LOCATION source. Use a small flat-blade screwdriver or similar The ORTEC 4002P Portable Power Supply is a tool to pop open the module door. Remove the completely portable, self-contained unit that is plastic drum: pinch the sides of the drum with your designed for table-top installation.
  • Page 11: Operating Instructions

    compatible with the principles of DOE Report TID- assured, at least with respect to the main connector 20893 (Rev), January 1968, "Standard Nuclear joining the module to the bin. Instrument Modules." This report, the work of a committee of equipment users from AEC-related The 4002P Portable Power Supply is a complete institutions, establishes standards for a modular assembly.
  • Page 12: Calibration And Maintenance

    terminals, the regulator automatically converts to a which receives its unregulated voltage from a current-foldback mode. This provides power limiting voltage-doubler circuit consisting of C1, C2, D1, and and protection of the regulator’s circuitry and components. When excess current demands are removed, the regulator resumes the voltage- Current flow to the output is monitored by sensing regulation mode.
  • Page 13: Factory Repair

    When the dc voltage levels, as monitored at the This instrument can be returned to the ORTEC front panel test points, are not within the prescribed factory for service and repair at a nominal cost. Our tolerances listed in Section 6.1, each can be standard procedure for repair ensure the same adjus ted.