Data Logger Programming; Kh20 Calibration; Maintenance And Calibration; Kh20 Calibration Ranges - Campbell KH20 Product Manual

Krypton hygrometer
Hide thumbs Also See for KH20:
Table of Contents

Advertisement

6.4

Data Logger Programming

6.4.1 KH20 Calibration

7.

Maintenance and Calibration

The KH20 sensor outputs 0 to 5 VDC analog signal. These signals can be
measured using the VoltDiff instruction on the CRBasic data loggers.
Programming basics for CRBasic data loggers are in the following sections.
Complete program examples for select CRBasic data loggers can be found in
Appendix B, Example Program
Each KH20 is calibrated over a vapor range of approximately 2 to 19 g/m
calibration is performed twice under the following two conditions: window
clean, and scaled. The water vapor absorption coefficient for three different
vapor ranges are calculated from the collected calibration data: full range, dry
range, and wet range. TABLE
over which three different water vapor absorption coefficients are calculated.
See Appendix A, Calibrating KH20
calibration.
TABLE 6-2. KH20 Calibration Ranges
Ranges
Full Vapor Range
Dry Vapor Range
Wet Vapor Range
Before the water vapor absorption coefficient, k
logger program for the KH20, the following decisions must be made:
Will the windows be allowed to scale?
What vapor range is appropriate for the site?
Once the decision is made, the appropriate k
calibrations sheet. The calibration sheet also contains the path length, x, for a
specific KH20. Using the water vapor absorption coefficient for either the dry
or the wet vapor range will produce more accurate measurements than using
that for the full range. If the vapor range of the site is unknown, or if the vapor
range is on the border line between the dry and the wet vapor ranges, the full
range should be used.
The KH20 sensor is designed for continuous field application and requires little
maintenance. The tube ends for the KH20 have been sealed with silicone
elastomer using an injection-mold method. Therefore, the tubes are protected
from water damage, and the KH20 continues to make measurements under
rainy or wet conditions. If the water tends to pool up on the tube window and
blocks the signal, turn the sensor head at an angle so as to shed the water off
the tube window. The rain shield protects the electronics box and the
connectors from moisture.
KH20 Krypton Hygrometer
.
(p. B-1)
6-2
shows a sample of the KH20 vapor ranges
, for more information on KH20
(p. A-1)
Vapor Density (g/m
2 – 19
2 – 9.5
8.25 – 19
, is entered into the data
w
can be chosen from the
w
3
. The
)
3
7