RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF
COMMUNICATIONS EQUIPMENT AND THE FARMADOLOR
The FarmaDolor is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the FarmaDolor can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the FarmaDolor as recommended
below, according to the maximum output power of the communications equipment.
RATED MAXIMUM OUTPUT
POWER OF TRANSMITTER
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in metres (m) can be estimated using the equation applicable to the frequency of
the transmitter, where P is the maximum output power rating of the transmitter in watts (W)
NOTE 1 - At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 - These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
SEPARATION DISTANCE ACCORDING TO FREQUENCY OF TRANSMITTER (M)
150 KHZ TO 80 MHZ
d = 1.2
0.12
0.38
1.2
3.8
12
according to the transmitter manufacturer.
80 MHZ TO 800 MHZ
800 MHZ TO 2.5 GHZ
d = 1.2
0.12
0.38
1.2
3.8
12
d = 2.3
0.23
0.73
2.3
7.3
23