Set-Point Value Is Not Reached (Mfc); Accessories, Spare Parts; Electrical Accessories - Burkert 8745 MFM Ethernet Operating Instructions Manual

Table of Contents

Advertisement

Type 8745 Ethernet, 8745 Analogue

Accessories, Spare Parts

12.3.12 Set-point value is not reached (MFC)

Identify the cause to solve the problem:
1. The mesh filter is clogged.
→ Clean or replace the mesh filter.
→ Run the Autotune function to adapt to the operating
conditions.
2. The inlet pressure is too low.
→ Increase the inlet pressure to the calibration pressure value.
3. The outlet pressure is too high.
→ If the fluid connection pipes after the product are dirty, then
clean them.
13
AccEssOriEs, spArE pArTs
CAUTION
risk of injury and risk of material damage that are due to
unsuitable parts.
Incorrect accessories and unsuitable replacement parts can
cause injuries and damage to the product and its environment.
▶ Only use original accessories and original spare parts from
Bürkert.
13.1

Electrical accessories

Tab. 36:
Electrical accessories, product variant Analogue with D-sub
DE-9 male connector
item
D-sub DE-9 female connector with 5 m
cable, with stripped end
D-sub DE-9 female connector with 10 m
cable, with stripped end
Tab. 37:
Electrical accessories, all product variants
item
büS stick set (without power supply)
Memory card
→ For further accessories, refer to the product datasheet.
English
article number
580 882
580 883
article number
772551
On request
65

Advertisement

Table of Contents
loading

This manual is also suitable for:

8745 mfm analogue8745 mfc ethernet8745 mfc analogue

Table of Contents