Main menu  Settings  Sensor settings  C1
Main menu  Settings  Sensor settings  C1
Further configuration possibilities regarding sensors, see Chapters 12.2.2.5 to 12.2.2.8!
See also Chapter 7.3.2.1.2.7 Textfelder beschriften und einstellen
Note:
After having confirmed by
accepted.
Caution:
Reference temperature and reference pressure (setting ex works 20°C, 1000 hPa):
All the volume flow (m³/h) and consumption values (m³) that are indicated on the display refer to 20°C
and 1000 hPa (according to ISO 1217 suction condition).
Alternatively, 0°C and 1013 hPa (=standard cubic meter according to DIN 1343) can also be entered
as the reference. Under no circumstances must the operating pressure or the operating temperature
be entered into the reference conditions!
®
METPOINT
BDL portable
OK,
the lettering is switched to black again. The values and settings have been
Operation METPOINT® BDL portable
Now, a
name
can be entered.
Subsequent to marking and confirming by OK,
the configuration of the sensor is completed.