REANIMATION WITH MASTER FLUX PLUS USING THE KIT Art. 1504/A "Kit for
reanimation with MASTER FLUX PLUS" (Not supplied):
(For further information see pict.35)
N.B.: Before carrying out the following operations set the percentage regulating
handle to 0. (0% Protoxide d'azoto)
1. Remove the Master Flux mask from the spiral pipe.
2. Insert the fitting (SMFA278) in the spiral pipe (CM88051).
3. Connect the extremity (A) of the pipe for oxygen therapy (CM84022) to the fitting
(SMFA278).
4. Connect the other extremity (B) of the pipe for oxygen therapy (CM84022) to the
stud of the reanimation balloon in position (C).
5. Connect the mask (CM84014) to the valve of the reanimation balloon (1025-2).
6. Turn the "FLOW" handle for supplying the oxygen flow you need, controlling the 0
column.
7. Put the oronasal mask (CM84014) on the patient.
8. Work on the balloon bag according to the instructions given for reanimation