12. The flux meter is equipped with a "FLUSH" push-
button POS.D (Cod.SMFA032) to be used in case
the patient shows malaise symptoms (nausea,
giddiness, headache) See PICT.2.
At the pressure of the "FLUSH" button the
bag (Cod.SMFA032) is filled with oxygen to
be manually delivered to the patient; if
necessary, interrupt the therapy. See PICT.2
13. Do not absolutely lubricate the medical equipment components.
14. The bottles used to supply a centralized system need to be located in an easily
accessible and dry place, having an antistatic floor and equipped with ventilating
holes directly connected to the outside.
15. The bottles have to be fixed to the wall with anti-turnover system (chain, etc...) and
equipped with a barrier (grid, etc...) for protecting them from impacts and
tampering.
16. Do not leave the bottles near heat sources (i.e. radiators, heating systems, flues,
etc...).
17. Do not install the bottles in hazardous places such as stores containing
inflammable substances, chemical compounds or in contact with electric
components.
18. Connect the pipes of the centralized system to the ground.
19. Do not absolutely lubricate any parts of the appliance.
20. Observations above described have to be carried out for assembling the centralized
system.
21. THE ASSEMBLING OF THE CENTRALIZED SYSTEM HAS TO BE CARRIED OUT BY A
SPECIALIZED COMPANY ENABLE TO RELEASE A CERTIFICATE DECLARING THE
SYSTEM HAS BEEN MANUFACTURED ACCORDING THE REGULATIONS IN FORCE
(with reference to the rules in force in countries where it is used).
22. Do not lean any object on the flow-meter box.
23. BEFORE CHARGING THE BOTTLES INDICATE ON THE RESPECTIVE COLLARS THE VAT
CODE OF THE OWNER (FOR THE ITALIAN MARKET ONLY).