The O
adpater port (with white plate) fits a hose to the end of which a male fitting is
2
stapled with 1/8 "G thread (Pict. 25 Pos. G.)
11. Connect the socket to the laboratory electric system equipotential point by means
of the supplied looped connectors (Sez. 2.5 mm
12. Connect the wall fixing plate connectors to the centralized system using nut and
union to be welded to the pipe by means of a copper Ø 10 mm (EN13348), see
pict.24 point C and D.
The picture below shows an example of installation of the centralised system.
PCT.25
BOTTLES
FIXATION
NOT ON
SUPPLY
A - Oxygen bottle
B - Protoxide bottle
C - Nitrous oxide reducer
D - Oxygen reducer
E - White tube
2
) see pict.24 point A and B.
PCT.24
N.B.: REDUCERS EQUIPPED WITH TUBES AND
COUPLES ARE INCLUDED IN 1501-AS
F - Blue tube
G - Male coupler 1/8" G
H - Male coupler ¼" G
I - Wall bracket
L - Soft copper pipes of the system