SPECIAL THERAPY FOR CHILDREN: (See pict.32 for any reference and position)
1. Find the proper flow the child needs, as before described.
2. To do this turn the handle "FLOW" (POS.A) clock-wise and let the patient inspire at
list for one minute.
3. Usually the flow needed by a child is around 4-5 Lt/min., see O
the flow-meter box and then the balloon (POS.B) that is to be half-full / half-empty.
4. Administer nitrogen protoxide (N
initially the indicator on the 20% and after few minutes watch the patient reactions.
5. If the patient has not sufficiently sedated improve the percentage for ex. 30% and so
on.
SPECIAL THERAPY FOR ADULTS: (See pict.32 for any reference and position)
1. Find the proper flow the adult patient needs, as before described.
2. To do this turn the handle "FLOW" (POS.A) clock-wise and let the patient inspire at
list for one minute.
3. Usually the flow for adult patients is around 6-7 Lt/min., see O
the flow-meter box and then the balloon (POS.B) that is to be half-full / half-empty.
4. Administer nitrogen protoxide (N
positioning first the indicator on the 20% and after few minutes watch the patient
reactions.
5. If the patient is not sufficiently sedated increase the percentage for ex. 30% and so
on.
END OF THERAPY: (See pict. 32 for reference and position)
1. At the end of the operation stop supplying Protoxide ( ) by turning the
handle "%" counter clock-wise up to its limit.
2. ADMINISTER, AS IN THE BEGINNING, OXYGEN (
Lt/min., FOR 2 MINUTES SETTING THE LITRES BY MEANS OF THE
HANDLE "FLOW" (SEE point. 6).
3. Stop supplying gas by turning the handle "FLOW" counter clock-wise up to its limit.
4. Remove the mask from the patient.
5. Close the tap of the protoxide (N
system the supply will be cut off at the end of the working day.
6. Close the tap of the Oxygen bottle (0
system the supply will be cut off at the end of the working day.
O) by means of the handle "%" (POS.C) positioning
2
O) by means of the handle "%" (POS.C) by
2
0) bottle (turn the tap clock-wise). For centralized
2
) (by turning clock-wise). For centralized
2
28
column (POS.D) on
2
column (POS.D) on
2
) ONLY TO (2÷3)