4. Take the flow-meter box (SMFA500) out of the packaging, screw the pin to the base
of the flow-meter box (SMFA313), (See pict.23).
5. Insert the pin (SMFA313) in the final part (POS.A) of the wall support (SMFA301) (See
pict.23).
6. Block the pin to the support (SMFA301) by means of the treated handle (CM15113)
(See pict.23).
7. Mount the nut (POS.B) of the Oxygen pipe (SMFA364 ) than connect to the O2
supplying nipple (POS.C) of the flow-meter box (SMFA500); mount the joint to the
other pipe extremity (POS.D) then connect to the check valve (POS.E) on the plate
(SMFA501). see pict.23
Mount the nut (POS.H) and the Nitrogen Protoxide pipe (SMFA365 ) to the supplying
8.
N2 O nipple (POS.G) of the flow-meter box (SMFA500); connect the joint to the other
pipe extremity (POS.H) than connect the check valve (POS.I) on the plate (SMFA501).
see pict.23.
PCT.23
9. Mount the 2 membrane reducers, that you find in the wall-type MASTER FLUX kit, on
the centralized system bottles.
10. The N
O adapter port (with blue plate) fits a hose to the end of which a male fitting
2
is stapled with ¼"G thread (Pict. 25 Pos. H).
N
O
2
VALVE
AFNOR
O
2
VALVE
AFNOR
23
ENGLISH
SMFA501
PCT.22