Dräger Evita
V-Series
®
Disinfection
Machine cleaning
Disinfection by immersion
Place the flow sensor in the disinfectant. Observe the specified contact time.
At the beginning of the contact time, swirl the flow sensor back and forth at least
3 times. Make sure the disinfectant reaches all surfaces and interiors.
At the end of the contact time, swirl the flow sensor back and forth again at least
3 times.
After the contact time, rinse the flow sensor in the water bath (at least drinking
water quality) until no more disinfectant residue can be detected.
Shake out residual water carefully. Allow the flow sensor to dry completely.
Check the flow sensor for visible damage, paying particular attention to its
6
measuring wires and their pins, and replace if necessary.
If you need visual descriptions, please look at the reprocessing instructions for manual cleaning.
Machine cleaning
Position the expiratory valve as shown. Position the components (flow sensor sleeve, diaphragm, water trap container)
to be stable.
Please ensure that all surfaces and interior spaces can be completely rinsed and the water can drain off freely.
This guideline does not replace or amend the instructions for use. Please always refer to the
reprocessing instruction in the latest instruction for use of the components. The instructions
for use and the hospital guidelines and policies always have to be observed.