Bowers M1 User Manual page 75

Display units for 1 or 2 air gauges
Table of Contents

Advertisement

M1 – M3 Display units
Instructions
RST
// Reset all the previous parameters.
1RES=4
// Change the resolution to display 4 digits (0.1µm)
1REF=PAIRING
PAIRING
1UT=+0000.0402
1NM=+0000.0000
1LT=+0000.0068
1MT=+0000.0068
1FM=5
// Formula = C1-C2
BAR=1
// Bar graph type = horizontal with left origin
DISPL=1
// Display type = simple (only 1 bar graph)
AGI
// Initialisation of air gauge calibration
AGL=1
// Configuration of input 1
AGF=0
// Filter level (from 0 to 5)
AGN=20.0000
// Value of the MIN master
AGA=20.0209
// Value of the MAX master
AGL=2
// Configuration of input 2
AGF=0
// Filter level (from 0 to 5)
AGN=19.9807
// Value of the MIN master
AGA=19.9932
// Value of the MAX master
AGS
// End of calibration configuration + display of the
configuration window
After this configuration, a popup window appears asking :
"DEACTIVATE CALIBRATION ?" YES / NO
Press yes and the previous calibration will be erased, then the window for calibrating the
display opens. You just need to finish the calibration procedure by placing the MIN and
MAX masters for C1 and C2.
Then go the measuring screen, place the master diam 20.0000 mm on C1 and diam
19.9932mm on C2, then press PRESET.
The value changes to 0.0068.
The display is ready to measure.
You can then scan a manufacturing order number if request to export the data, as
describes on the chapter "DATA EXPORT ON USB KEYS".