Calibration In 3 Points - Bowers M1 User Manual

Display units for 1 or 2 air gauges
Table of Contents

Advertisement

M1 – M3 Display units
After calibration is it possible to check the pressure. For this you need to click on the
unit "mm", the unit will then become BAR for few seconds.
This feature allows for example to check if the restrictor installed on the air input is
the right one. The ideal pressure at the nominal should be 2.7 BAR. If you are too far
from this value, for example 1.5 BAR, you should change the restrictor for a bigger
one. Using the wrong restrictor affects the linearity. For a simple sorting Go/NoGo,
the results will be correct, but if you need to have the real value and/or to use the
class sorting functionality, you need to use the most adapted restrictor. (see chap
3.3.1).
6.3.1.2.

Calibration in 3 points

There is the possibility to use 3 points to calibrate the measure. When the button
"central" is selected, a third master point is required. The calibration process stays the
same, with one extra calibration point added.
This feature can be useful when controlling a part with large tolerances.
NB. Calibration sizes must be taken from calibration certificate.