11. CONFIGURATION OF THE UNIT WITH QR CODES

The M1/M3 displays offers the possibility to be configured by a simple scan of QR
code. You should use the QR Code reader 55-QR1
This will simplify the configuration and reduce risks of errors. It takes less than 2
seconds to configure the display with a QR code scan.
List of commands :
All the RS232 commands listed on the table chapter 10 - RS232 communication can
be encoded on a QR code.
Additionally, to this, specifics commands are reserved for the calibration procedure of
the M1/M3.

11.1. Calibration commands:

It is possible to launch a calibration procedure with the QR code:
The QR code must be encoded according to the following layout:
Instructions
AGI
// Initialisation
AGL=1
// Configuration of input 1
AGF=1
// Filter level (from 0 to 5)
AGN=20.00000
// Value of the MIN master
AGC=20.01000
// value of the CENTRAL master (optional)
AGA=20.02090
// Value of the MAX master
AGS
// End of calibration configuration + display of the
configuration window
If you want to configure 2 channels, same principle but you should insert the
commands to calibrate the channel 2 :
Instructions
AGI
// Initialisation
AGL=1
// Configuration of input 1
AGF=1
// Filter level (from 0 to 5)
AGN=20.00000
// Value of the MIN master
AGC=20.01000
// value of the CENTRAL master (optional)
AGA=20.02090
// Value of the MAX master
AGL=2
// Configuration of input 2
AGF=2
// Filter level (from 0 to 5)
AGN=19.9807
// Value of the MIN master
AGA=19.9932
// Value of the MAX master
AGS
// End of calibration configuration + display of the
configuration window
Corresponding QR
Code
Corresponding QR
Code
Page 73