Safety Information
Intended Use
Intended Application of the Gas Analyzer
Safety Instructions
Requirements for Safe Operation
Qualifications of the Personnel
The gas analyzers of the EL3060 series are intended for the continuous
quantitative determination of individual gas components in gas mixtures.
Any other application is not compliant with the specified use. The
specified use also includes taking note of this operator's manual.
The gas analyzer is suitable for measuring non-flammable and flammable
gases under atmospheric conditions, which can occasionally form a
potentially explosive atmosphere (Zone 1). The mixing ratio of these
gases should be clearly below the lower explosive limit (LEL) or clearly
above the upper explosive limit (UEL). Exceptions can be e.g. startup and
shutdown conditions.
A special version of the gas analyzer is suitable for measuring non-
flammable and flammable gases under positive pressure (see page 20) if
special conditions are met.
The gas analyzer may not be used for measuring gases that attack the
materials in contact with the sample medium (e.g. gases containing
chlorine).
The EL3060-Uras26 analyzer unit may only be operated in conjunction
with the EL3060-... control unit.
In order to operate in a safe and efficient manner the device should be
properly handled and stored, correctly installed and set up, properly
operated and correctly maintained.
Only persons familiar with the installation, commissioning, operation and
maintenance of comparable devices and certified as being capable of
such work should work on the device.
Safety Information
7