Pressure Sensor - ABB EasyLine EL3060 Series Instructions For Installation Start-Up And Operation

Continuous gas analyzers for use in hazardous locations
Hide thumbs Also See for EasyLine EL3060 Series:
Table of Contents

Advertisement

24
EasyLine EL3060 Series Commissioning Instructions

Pressure Sensor

Uras26
Magnos206
Caldos27
Caldos25
Information for the safe and correct operation of the pressure sensor
The pressure sensor is installed in the gas analyzer as standard. It is
optionally connected to a gas port via an FPM tube (flame barrier).
The pressure sensor is optionally installed in the gas analyzer. It is
optionally connected to a gas port via an FPM tube (flame barrier).
The pressure sensor is installed in the gas analyzer as standard. It is
optionally connected to a gas port via an FPM tube (flame barrier).
No pressure sensor is installed in the gas analyzer.
·
If the pressure sensor (see page 24) is connected to the outside by
hose, the yellow plastic screw cap must be screwed out of the
connection fittings of the pressure sensor (flame barrier) before the
gas analyzer is commissioned.
·
If the pressure sensor connection is not connected to the sample gas
outlet, an exact pressure correction is required so that the pressure
sensor and the sample gas outlet are on the same pressure level.
ATTENTION
The pressure sensor connection must not be connected to the sample
gas feed path for the measurement of flammable gases.