EasyLine EL3060 Series Commissioning Instructions
·
Do the 24 VDC connecting cable and the data transmission cable
which are securely connected to the EL3060-Uras26 analyzer unit
have a length of more than 1 m and do they have any damage?
Integrity of the Case of the EL3060-Uras26 Analyzer Unit
·
Is the case of the analyzer unit intact?
·
Are all flame barriers and screw plugs present?
·
In the case of horizontal assembly of the analyzer unit: Are the O-
rings, which have been inserted in the grooves provided between the
bottom of the case and the case and between the case and the case
cover, clean and not pinched?
·
Have all components of the case been completely bolted together
and locked against rotation with the socket-head hex screws?
Integrity of the Case of the Control Unit
·
Is the case of the control unit intact?
·
Is the case of the control unit tightly closed?
·
Is the case cover screwed in all the way and locked against rotation
with the socket-head hex screw?
·
Is the seal in the cover of the terminal housing intact? Is the cover of
the terminal housing tightly closed?
·
Are all screwed cable glands present and tightly screwed in?
·
Are the openings of the unused screwed cable glands tightly closed
with sealing plugs?
Connection of the Peripheral Units
·
Are all devices needed for gas conditioning, calibration and waste
gas disposal correctly connected and ready for use?