Cannula settings
How to set cannula and scope combination or Dynamic Mode settings.
Step 1: Access settings menu
= Access settings menu
Note: The selected Cannula setting is displayed
on the top banner of the Run Screen.
The IFU part number is: P17330. The CrossFlow Pump catalog number is: 0450000000
Step 2: Choose a desired option
Dynamic Mode options provide optimized flow for a range of scope cannula combinations.
Hardware-specific provides customized flow for the scope cannula combination you are using.
While operating in "Dynamic Mode", select one of the following options:
Dynamic mode setting
Cannula diameter (mm) – scope diameter (mm) = x
Dynamic Low
x ≤ 1mm
Dynamic Medium
1 mm < x ≤ 2mm
Dynamic High
x > 2mm
Dynamic Mode
Select a cannula setting
= Cannula settings menu
Hardware-specific alternative tab
As an alternative to Dynamic Mode cannula
settings, the specific scope/cannula combination
can be selected in the hardware-specific tab.