Stryker IsoTour Maintenance Manual

Hide thumbs Also See for IsoTour:

Advertisement

Quick Links

I I s s o o T T o o u u r r ™ ™ P P u u m m p p
M M a a i i n n t t e e n n a a n n c c e e M M a a n n u u a a l l
2874
EN
2874-009-002 Rev A.1
2019/08

Advertisement

loading

Summary of Contents for Stryker IsoTour

  • Page 1 I I s s o o T T o o u u r r ™ ™ P P u u m m p p M M a a i i n n t t e e n n a a n n c c e e M M a a n n u u a a l l 2874 2874-009-002 Rev A.1 2019/08...
  • Page 3 S S y y m m b b o o l l s s Refer to instruction manual/booklet Operating instructions/Consult instructions for use General warning Caution Cardiopulmonary resuscitation (CPR), disconnection point Catalogue number Serial number Batch code Manufacturer Authorized representative in the European Community European medical device CE mark 2797...
  • Page 4 Solids: Protection from touch by fingers and objects greater than 12 mm I I P P 2 2 1 1 Liquids: Protection from the fall of vertical water drops For US Patents see www.stryker.com/patents Keep dry Do not stack more than 6 high...
  • Page 5 T T a a b b l l e e o o f f C C o o n n t t e e n n t t s s Warning/Caution/Note Definition ..........................3 Summary of safety precautions ..........................3 Introduction ................................4 Product description ..............................4 Intended use ................................4 Expected service life ...............................4...
  • Page 6: Emc Information

    EMC information ..............................35 2874-009-002 Rev A.1...
  • Page 7 W W a a r r n n i i n n g g / / C C a a u u t t i i o o n n / / N N o o t t e e D D e e f f i i n n i i t t i i o o n n The words W W A A R R N N I I N N G G , C C A A U U T T I I O O N N , and N N O O T T E E carry special meanings and should be carefully reviewed.
  • Page 8 The Stryker Model 2874, I I s s o o T T o o u u r r ™ ™...
  • Page 9 N N o o t t e e - - T T r r u u T T u u r r n n requires the I I s s o o T T o o u u r r pump to be connected for use. The T T r r u u T T u u r r n n feature is visually identified by the icons on the I I s s o o T T o o u u r r pump (Figure 1).
  • Page 10 Competent authority of the European Member State where the user and/or patient is established. To view your operations or maintenance manual online, see https://techweb.stryker.com/. Have the serial number (A) of your Stryker product available when calling Stryker Customer Service or Technical Support. Include the serial number in all written communication.
  • Page 11 S S e e r r i i a a l l n n u u m m b b e e r r D D a a t t e e o o f f m m a a n n u u f f a a c c t t u u r r e e The year of manufacture is the first four digits of the serial number.
  • Page 12 • Do not place unprotected circuit boards on the floor. N N o o t t e e - - Always ship back circuit boards to Stryker in the same antistatic bags that the new boards were originally shipped in.
  • Page 13 P P r r o o c c e e d d u u r r e e : : 1. Unplug the pump from AC power. 2. Unplug the mattress tubing from the pump. 3. Place the pump with the front pump housing side down on a work surface. 4.
  • Page 14: Phillips Screwdriver

    L L C C D D s s c c r r e e e e n n r r e e p p l l a a c c e e m m e e n n t t T T o o o o l l s s r r e e q q u u i i r r e e d d : : •...
  • Page 15 13. Verify proper operation before you return the product to service. H H o o o o k k r r e e p p l l a a c c e e m m e e n n t t T T o o o o l l s s r r e e q q u u i i r r e e d d : : •...
  • Page 16 6. Separate the front and back pump housing. Place the back pump housing down on the work surface. C C A A U U T T I I O O N N - - Always use a grounded static strap to prevent static coming into contact with the PCB assembly. 7.
  • Page 17 P P o o w w e e r r i i n n l l e e t t r r e e p p l l a a c c e e m m e e n n t t T T o o o o l l s s r r e e q q u u i i r r e e d d : : •...
  • Page 18 11. Remove and discard the PCB assembly. 12. Reverse the steps to reinstall ( PCBA assembly 2874-007-025 (page 27)). 13. Run and pass the diagnostic test. 14. Verify proper operation before you return the product to service. L L E E D D b b o o a a r r d d r r e e p p l l a a c c e e m m e e n n t t T T o o o o l l s s r r e e q q u u i i r r e e d d : : •...
  • Page 19 6. Separate the front and back pump housing. Place the back pump housing down on the work surface. 7. On the solenoid/valve that needs to be replaced, make note of the hose locations first, then unplug each hose from the valve.
  • Page 20 A A c c c c e e s s s s d d i i a a g g n n o o s s t t i i c c m m o o d d e e P P r r o o c c e e d d u u r r e e : : 1.
  • Page 21 N N o o t t e e • At a minimum, check all items listed during annual preventive maintenance for all Stryker Medical products. You may need to perform preventive maintenance checks more often based on your level of product usage.
  • Page 22 T T r r o o u u b b l l e e s s h h o o o o t t i i n n g g P P r r o o b b l l e e m m S S c c r r e e e e n n C C a a u u s s e e R R e e c c o o m m m m e e n n d d e e d d a a c c t t i i o o n n...
  • Page 23 P P o o w w e e r r c c o o r r d d s s P P a a r r t t N N u u m m b b e e r r Power cord, type B, 1 meter 2874-007-001 Power cord, type B, 5 meter...
  • Page 24 P P u u m m p p h h o o u u s s i i n n g g a a s s s s e e m m b b l l y y 2 2 8 8 7 7 4 4 - - 0 0 0 0 7 7 - - 0 0 1 1 7 7 2874-009-002 Rev A.1...
  • Page 25 T T o o p p p p a a n n e e l l a a s s s s e e m m b b l l y y 2 2 8 8 7 7 4 4 - - 0 0 0 0 7 7 - - 0 0 1 1 8 8 I I t t e e m m N N u u m m b b e e r r N N a a m m e e...
  • Page 26 H H o o o o k k a a s s s s e e m m b b l l y y 2 2 8 8 7 7 4 4 - - 0 0 0 0 7 7 - - 0 0 1 1 9 9 I I t t e e m m N N u u m m b b e e r r N N a a m m e e...
  • Page 27 B B u u m m p p e e r r p p a a c c k k 2 2 8 8 7 7 4 4 - - 0 0 0 0 7 7 - - 0 0 2 2 0 0 I I t t e e m m N N u u m m b b e e r r N N a a m m e e...
  • Page 28 P P u u m m p p h h o o u u s s i i n n g g r r u u b b b b e e r r f f o o o o t t a a s s s s e e m m b b l l y y 2 2 8 8 7 7 4 4 - - 0 0 0 0 7 7 - - 0 0 2 2 2 2 I I t t e e m m N N u u m m b b e e r r N N a a m m e e...
  • Page 29 C C o o m m p p r r e e s s s s o o r r a a s s s s e e m m b b l l y y 2 2 8 8 7 7 4 4 - - 0 0 0 0 7 7 - - 0 0 2 2 3 3 I I t t e e m m N N u u m m b b e e r r N N a a m m e e...
  • Page 30 P P o o w w e e r r s s u u p p p p l l y y a a s s s s e e m m b b l l y y 2 2 8 8 7 7 4 4 - - 0 0 0 0 7 7 - - 0 0 2 2 4 4 I I t t e e m m N N u u m m b b e e r r N N a a m m e e...
  • Page 31 P P C C B B A A a a s s s s e e m m b b l l y y 2 2 8 8 7 7 4 4 - - 0 0 0 0 7 7 - - 0 0 2 2 5 5 I I t t e e m m N N u u m m b b e e r r N N a a m m e e...
  • Page 32: Hepa Filter

    H H E E P P A A f f i i l l t t e e r r 2 2 8 8 7 7 4 4 - - 0 0 0 0 7 7 - - 0 0 2 2 6 6 I I t t e e m m N N u u m m b b e e r r N N a a m m e e...
  • Page 33 A A i i r r f f i i l l t t e e r r g g u u a a r r d d 2 2 8 8 7 7 4 4 - - 0 0 0 0 7 7 - - 0 0 2 2 7 7 I I t t e e m m N N u u m m b b e e r r N N a a m m e e...
  • Page 34 3 3 w w a a y y d d i i g g i i t t a a l l v v a a l l v v e e a a s s s s e e m m b b l l y y 2 2 8 8 7 7 4 4 - - 0 0 0 0 7 7 - - 0 0 2 2 8 8 I I t t e e m m N N u u m m b b e e r r N N a a m m e e...
  • Page 35 5 5 w w a a y y d d i i g g i i t t a a l l v v a a l l v v e e a a s s s s e e m m b b l l y y 2 2 8 8 7 7 4 4 - - 0 0 0 0 7 7 - - 0 0 2 2 9 9 I I t t e e m m N N u u m m b b e e r r N N a a m m e e...
  • Page 36 F F r r o o n n t t L L E E D D m m o o d d u u l l e e a a s s s s e e m m b b l l y y 2 2 8 8 7 7 4 4 - - 0 0 0 0 7 7 - - 0 0 3 3 0 0 I I t t e e m m N N u u m m b b e e r r N N a a m m e e...
  • Page 37 H H a a n n d d l l e e a a s s s s e e m m b b l l y y 2 2 8 8 7 7 4 4 - - 0 0 0 0 7 7 - - 0 0 3 3 1 1 I I t t e e m m N N u u m m b b e e r r N N a a m m e e...
  • Page 38 M M a a t t t t r r e e s s s s h h o o s s e e a a s s s s e e m m b b l l y y 2 2 8 8 7 7 4 4 - - 0 0 0 0 7 7 - - 0 0 3 3 4 4 518M104001 2874-009-002 Rev A.1...
  • Page 39 E E M M C C i i n n f f o o r r m m a a t t i i o o n n W W A A R R N N I I N N G G - - The use of accessories, transducers, and cables, other than those specified or provided by the manufacturer, could result in increased electromagnetic emissions or decreased electromagnetic immunity and result in improper operation.
  • Page 40 R R e e c c o o m m m m e e n n d d e e d d s s e e p p a a r r a a t t i i o o n n s s d d i i s s t t a a n n c c e e s s b b e e t t w w e e e e n n p p o o r r t t a a b b l l e e a a n n d d m m o o b b i i l l e e R R F F c c o o m m m m u u n n i i c c a a t t i i o o n n e e q q u u i i p p m m e e n n t t a a n n d d t t h h e e 2 2 8 8 7 7 4 4 I I s s o o T T o o u u r r p p u u m m p p 380-390 TETRA 400...
  • Page 41 G G u u i i d d a a n n c c e e a a n n d d m m a a n n u u f f a a c c t t u u r r e e r r ’ ’ s s d d e e c c l l a a r r a a t t i i o o n n - - e e l l e e c c t t r r o o m m a a g g n n e e t t i i c c i i m m m m u u n n i i t t y y Main power quality should be that of a typical for 0.5 cycle at 0°,...
  • Page 42 Portable and mobile RF communications equipment should be used no closer to any part of the 2874 I I s s o o T T o o u u r r pump, including cables, than the recommended separation distance calculated from the equation appropriate for the frequency of the transmitter.
  • Page 44 Stryker Medical 3800 E. Centre Avenue Portage, MI 49002 2874-009-002 Rev A.1 2019/08 WCR: AA.21...