Stryker Care-Free Companion CF300 Functional Check Instructions
Stryker Care-Free Companion CF300 Functional Check Instructions

Stryker Care-Free Companion CF300 Functional Check Instructions

Advertisement

Quick Links

Care•Free Companion
FUNCTIONAL CHECK INSTRUCTIONS
Read and understand all instructions and the following safety precautions before beginning any work.
SAFETY PRECAUTIONS
DANGER
• Electrical shock hazard. Do not remove cover; there are no serviceable parts inside.
• Explosive hazard. Do not use in the presence of flammable anesthetics.
WARNING
• Disinfect before use with each patient, utilizing standard hospital protocol and disinfectants. Failure to disinfect may
cause cross-contamination and infection.
• Do not autoclave. Autoclaving will damage the pump.
CAUTION
• The Functional Check is an optional test, however should you choose to perform the test it must be performed only
by qualified technical personnel familiar with good practice for medical device repair.
• Verify proper operation of the pump. Operation of the pump at less than specified flows and pressures may result
in poor or reduced patient therapy.
Warning - Consult
accompanying documentation
Caution
Shock Hazard
TOOLS REQUIRED
Product
Tool
CF300
Tester
Voltage: 120V
Functional Check Instructions
Frequency: 60Hz
Manometer
Tubing
GENERAL DESCRIPTION
The Functional Check Instructions provide a quantitative means of determining whether a Care•Free Companion
Pump can properly inflate its cushion. An alternate, qualitative means to check pump and cushion performance is to
follow the HAND CHECK described in the Care•Free Companion Pump Instructions for Use.
The Functional Check Instructions require the use of a Tester tool that simulates the cushion that is normally used with
these pumps. If the pump does not meet the outlined requirement, it should be replaced. Repair is not available for
these pumps.
PHYSICAL INSPECTION
1. Unplug the power cord.
2. Check tubing and cap for visible signs of damage (kinks, cuts, and cracks).
2012/12 Rev C.0
®
Pump
CF300
100259 Rev C
Manufacturer
Model
Do not use with Flammable Anesthetics
Description
P/N 12157
P/N 100259
0-50 mm Hg or greater (standard mercury, or digital)
3/16" ID
Page 1 of 2

Advertisement

loading

Summary of Contents for Stryker Care-Free Companion CF300

  • Page 1 Care•Free Companion ® Pump CF300 FUNCTIONAL CHECK INSTRUCTIONS Read and understand all instructions and the following safety precautions before beginning any work. SAFETY PRECAUTIONS DANGER • Electrical shock hazard. Do not remove cover; there are no serviceable parts inside. • Explosive hazard. Do not use in the presence of flammable anesthetics. WARNING • Disinfect before use with each patient, utilizing standard hospital protocol and disinfectants.
  • Page 2 4. With the finger removed from the orifice, the pressure reading on the manometer should be at least 13 mm Hg. If any of the pressure readings are out of specification, discontinue use and contact your local dealer or Stryker’s Customer Service.