Service Manual
836004-5210 V1.1
2870
REF
2871
REF
www.stryker.com

Advertisement

Table of Contents
loading

Summary of Contents for Stryker EOLE

  • Page 1 EOLE / EOLE DC Pump 2870 2871 Service Manual 2017/10 A.0 www.stryker.com 836004-5210 V1.1...
  • Page 3: Table Of Contents

    Air cell replacement ....................................5 Check internal tubing ....................................5 Fuse replacement ......................................6 Main PCBA replacement ..................................6 Timing motor module replacement ..............................8 Micro switch replacement ..................................9 Front Panel PCBA replacement ................................10 Filter replacement ....................................10 Pressure Calibration ....................................11 Pressure test ......................................13 www.stryker.com 836004-5210 V1.1...
  • Page 4: Troubleshooting

    Switch assembly, with PVC switch cap 2870001010 EOLE DC/EOLE Front case enclosure 2870001012 EOLE DC/EOLE Mattress bottom cover 32” (80cm) 2871019009 EOLE DC Back Case 2870001013 EOLE/EOLE DC Pump Service Manual for Eole and Eole DC 2870009002 EOLE/EOLE DC 836004-5210 V1.1 www.stryker.com...
  • Page 5: Service

    Remove wire connections to power switch and front panel PCBA. Take out the module set from the case. Pumping set is inside the compressor. Unscrew the 4 screws on the compressor and open the case. Remove the screw of pumping set. Take out the pumping set. www.stryker.com 836004-5210 V1.1...
  • Page 6 Use your thumb to push back the pumping set and hold on this position, and then tighten the screw. Press the vibration rods to make sure they will not touch the front object. Apply threadlocker around the screw to set for the best bond. 836004-5210 V1.1 www.stryker.com...
  • Page 7: Air Cell Replacement

    Remove the 8 rubber stops. Using a Philips screwdriver, remove the 8 screws that hold the rear case. Remove the rear case. Inspect for the tubing connections. Make sure that each tube is attached to its connector. www.stryker.com 836004-5210 V1.1...
  • Page 8: Fuse Replacement

    TOOLS REQUIRED:  ● Phillips screwdriver for M3 screws PROCEDURE: Remove the 8 rubber stops. Using a Philips screwdriver, remove the 8 screws that hold the rear case. Remove the rear case. Remove 6 screws on the base. 836004-5210 V1.1 www.stryker.com...
  • Page 9 Remove the 2 screws and all wire connections on the main PCBA. Take out the main PCBA. Replace with a new one and discard the old one. The wire connections on the main PCBA are as below. After replacement, perform pressure calibartion. Refer to Pressure Calibration section. www.stryker.com 836004-5210 V1.1...
  • Page 10: Timing Motor Module Replacement

    Remove the tubes and wire connections from timing motor module. Remove the 2 screws on the motor assembly. Remove the wire of timing motor module from the main PCBA. Take out the timing motor module. Replace with a new one and discard the old one. 836004-5210 V1.1 www.stryker.com...
  • Page 11: Micro Switch Replacement

    Remove the 8 rubber stops. Using a Philips screwdriver, remove the 8 screws that hold the rear case. Remove the rear case. Remove the wire connections from micro switch. Remove the 2 screws on the micro switch and then take out the micro switch. Replace with a new one and discard the old one. www.stryker.com 836004-5210 V1.1...
  • Page 12: Front Panel Pcba Replacement

    Remove the tubes and wire connections from front panel PCBA. Remove the 5 screws on the front panel PCBA. FILTER REPLACEMENT PROCEDURE: Open the filter case. Take out the filter. Replace with a new one and discard the old one. 836004-5210 V1.1 www.stryker.com...
  • Page 13: Pressure Calibration

    Press “A”. There will be a long beep. After 10 seconds, press “A” again. There will be a long beep. Remove the sensor tube from the main PCBA. Connect the tubing of the calibration tool onto the sensor on the main PCBA. www.stryker.com 836004-5210 V1.1...
  • Page 14 13. Remove the calibration tubing and put the sensor tube on senor. 14. Close the rear case and tighten all screws. Note: Once you displace the main PCBA, you must complete the pressure calibration and pressure test. 836004-5210 V1.1 www.stryker.com...
  • Page 15: Pressure Test

    Choose leve1, and after the pressure is stable, check the pressure meter. The pressure should be 20 ± 3 mmHg. Choose level 5, and after the pressure is stable, check the pressure meter. The pressure should be 32 ± 5 mmHg. Note: If the pressure is out of range, please repeat the calibration process. www.stryker.com 836004-5210 V1.1...
  • Page 16 Stryker Medical 3800 E. Centre Avenue Portage, Michigan 49002 Stryker European Operations B.V. Herikerbergweg 110 Amsterdam 1101 CM Netherlands 2017/10 www.stryker.com 836004-5210 V1.1...