Current meter
Table of Contents

Advertisement

Model 801 EM Current Meter
Page 14
# I D E N T
S P E E D
S D =
A V
P E R I O D
4.6.3. Extracting Data
Selecting EXTRACT DATA at the LOGGING MENU reveals the following display.
Connect the unit to a PC via the data interface lead supplied. Run a terminal emulation program
on the PC, ensuring that communications are correctly set to 4800 baud, 8 data bits, 1 stop bit, no
parity bits, flow control NONE. If the data is to be saved on the PC, make sure that the data is
directed to a file name. It is uploaded as a text file, with "Tab" delimiters, so it can be read into a
word processor or spreadsheet application.
The data lead connector information is:
CDU end
4 way in-line Male MilSpec connector [LMH06F 08
04 PN] [pins]
UPLOAD
Begins to upload data to PC. Screens similar to those shown below will appear, and
during uploading the #IDENT will increment.
EXIT
Returns to LOGGING MENU. Refer to Section 4.6
© 2016 Valeport Ltd
F F F R
+ X . X X X
X . X X X
S E C S
S S S
E X T R A C T
P L E A S E
C O N N E C T
Pin A
Pin B
Pin C
Pin D
D A T A
P C
Function
RTS from PC [not
used]
Tx RS232 from PC
to CDU
Gnd
Rx RS232C to PC
E X I T > > >
U P L O A D > > >
E X I T > > >
PC end
9 way "D" type
female [sockets]
Socket 7
Socket 3
Socket 5
Socket 2