Table 1: Technical Data:
Manufacturer
Type
Dimensions
Width
Depth
Height
Weight with rotor
max. speed
max. volume
max. RCF
allowable density
allowable kinetic energy
Mains power connection AC
Voltage fluctation
Current consumption
Power consumption
Audit requirement (BGR 500)
Ambient conditions
- Environment
- High
- Ambient temperature
- Max. relative humidity
- Overvoltage category (IEC 60364-4-443)
- Degree of contamination
Class of protection I
Noise level (depending on the rotor)
Write from operator:
Inventory-No.:
Monitoring-No.:
Environement:
Maintenance contact:
responsible service office
responsible dealer
I
Benchmark Scientific inc.
Z 130 M
14,6 cm
20,3 cm
10,8 cm
1,3 kg
5500 min
12 x 1.5/2.0 ml and 4 x PCR-stripes
2000 x g
1,2 kg/dm
58 Nm
230 V / 50-60 Hz 1 ph
0,12 A
0,04 KW
nein
(EN / IEC 61010-1)
Max. relative humidity 80 % for temperatures up to 31°C,
Decreasing linearly to 50 % relative humidity up to 35°C.
Not suitable for use in hazardous environments
HERMLE Labortechnik GmbH
Siemensstraße 25
78564 Wehingen
Tel.: (49)7426 / 96 22-17
Fax: (49)7426 / 96 22-49
Z 130 M_V1.15
-1
3
120 V / 50-60 Hz 1 ph
± 10 %
0,18 A
0,04 KW
for indoor use only
Use up to an altitude of 2000 m above MSL
2°C up to 35 °C
II
2
Class of protection (DIN EN 60529) IP 20
52 +2,0 dB(A)